User Contributed 7411B PROCEDURE ================================================================================ INSTRUMENT: H - P 5314A Universal Counter No Options. DATE: 21-Oct-88 AUTHOR: User Contributed REVISION: 0.1 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 26 NUMBER OF LINES: 119 CONFIGURATION: Fluke 6060B Wavetek 278A =============================================================================== STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DRAW 6 78 7 78 80 5 60 45 1.002 ASK- J 1.003 DISP Plug the unit in & let it warm up 1/2 hour if you have 1.003 DISP not already done so. [DRAW20,5,F] 1.004 HEAD {Time Base Check & Adjustment.} 1.005 SET ON/OFF: ON [DRAW20,5,F] 1.006 SET NORM/HOLD: NORM [DRAW20,10,F] 1.007 SET FUNCTION: FREQ A [DRAW20,20,F] 1.007 SET BLUE Key: OUT 1.008 SET RESOLUTION: 1 Hz [DRAW20,30,F] 1.009 SET CHANNEL A LEVEL A: Centered [DRAW60,60,F] 1.009 SET CHANNEL A SLOPE: + 1.009 SET CHANNEL A ATTN: X1 1.010 SET CHANNEL B LEVEL B: Centered [DRAW60,90,F] 1.010 SET CHANNEL B SLOPE: + 1.011 SET SEP/COM A: SEP [DRAW40,80,F] 1.012 MESS Adjust LEVEL control(s) as necessary. 1.013 DISP Use INPUT A. [DRAW20,60,F] 1.014 6060 9900KH .04U .1V -D 1.015 6060 9900KH .1V S 1.016 DISP Adjust OSC ADJ for 9.9 MHz +/- 40 Hz. [DRAW0,5,F] 2.001 HEAD {Frequency Response & Sensitivity.} 2.002 M278 B4 2.003 278 .011KH .001U 25mV -D SI L 3.001 278 .1KH .001U 25mV -D SI L 4.001 278 1KH .001U 25mV -D SI L 5.001 278 10KH .001U 25mV -D SI L 6.001 278 100KH .0014U 25mV -D SI L 7.001 6060 1000KH .004U 25mV -D 8.001 SET RESOLUTION: 10Hz N=1 [DRAW20,30,F] 8.002 6060 10000KH .05U 25mV -D 9.001 6060 30000KH .13U 25mV -D 10.001 6060 70000KH .29U 25mV -D 11.001 6060 99000KH .41U 25mV -D 12.001 SET SEP/COM A: COM A [DRAW40,80,F] 12.002 SET FUNCTION: All 3 Switches Out [DRAW20,20,F] 12.003 SET RESOLUTION: 1 Hz [DRAW20,30,F] 12.004 278 10H 25mV SI S L 12.005 EVAL {Does the display read 1.00?} 13.001 278 1KH 25mV SI S L 13.002 EVAL {Does the display read 1.00?} 14.001 278 100KH 25mV SI S L 14.002 EVAL {Does the display read 1.00?} 15.001 278 2.5MH 25mV SI S L 15.002 EVAL {Does the display resd 1.00?} 16.001 HEAD {Period.} 16.002 SET FUNCTION: PER A [DRAW20,20,F] 16.003 SET RESOLUTION: 10 Hz N = 1 [DRAW20,30,F] 16.004 SET SEP/COM A: SEP [DRAW40,80,F] 16.005 278 90000uT 2.5V SI S L 16.006 EVAL {Does the display indicate a period near 90000 uS?} 17.001 SET RESOLUTION: 1 Hz [DRAW20,30,F] 17.002 278 1000uT .001U .0014% 250mV -D SI L 18.001 278 .4uT .001U .0014% 25mV -D SI L 19.001 HEAD {Time Interval.} 19.002 SET FUNCTION: TI A --> B [DRAW20,20,F] 19.002 SET BLUE Key: IN 19.003 SET RESOLUTION: N = 1 [DRAW20,30,F] 19.004 SET CHANNEL B SLOPE: - [DRAW40,90,F] 19.005 SET SEP/COM A: COM A [DRAW40,80,F] 19.006 278 1MH .1V SI S L 19.007 EVAL {Does the display read 0.5 uS +/- 0.1 uS?} 20.001 HEAD {Ratio.} 20.002 SET FUNCTION: RATIO A/B [DRAW20,20,F] 20.002 SET BLUE Key: OUT 20.003 SET RESOLUTION: 1 Hz [DRAW20,30,F] 20.004 SET CHANNEL B SLOPE: + [DRAW40,90,F] 20.005 SET SEP/COM A: SEP [DRAW40,80,F] 20.006 ASK- W 20.007 DISP Connect FUNC OUT to INPUT B. [DRAW20,90,F] 20.008 DISP Connect SIGNAL GENERATOR to INPUT A. [DRAW20,60,F] 20.009 278 2MH 25mV SI S L 20.010 6060 10MH 25mV S 20.011 EVAL {Does the display read 5.00?} 21.001 ASK+ W 21.002 HEAD {Totalize.} 21.003 SET FUNCTION: START A [DRAW20,20,F] 21.003 SET BLUE Key: IN 21.004 DISP Connect FUNC OUT to INPUT A. [DRAW20,60,F] 21.005 278 10H 25mV SI S L 21.006 EVAL {Is the counter counting upward at a 10 Hz rate?} 22.001 SET NORM/HOLD: HOLD [DRAW20,10,F] 22.002 278 10H 25mV SI S L 22.003 EVAL {Did the display STOP counting & is it HOLDING its display?} 23.001 SET NORM/HOLD: NORM [DRAW20,10,F] 23.002 278 10H 25mV SI S L 23.003 EVAL {Has the counter resumed counting & is it STILL counting?} 24.001 278 10MH 25mV SI S L 24.002 EVAL {Is the counter STILL counting?} 25.001 278 10MH 25mV SI S L 25.002 EVAL {Is the OVFL indicator lit?} [DRAW60,5,F] 26.001 DISP W 26.001 DISP O 26.001 DISP W 26.001 DISP ! 26.002 278 * S 26.003 6060 * S 26.004 M278 B0 END